联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

When running SPC (signal path compensation) on my oscilloscope, I get the message that SPC failed.  What does failing SPC indicate?

问题:

When running SPC (signal path compensation) on my oscilloscope, I get the message that SPC failed.  What does failing SPC indicate?

答案:

When an oscilloscope says SPC failed, 99.9 percent of the time it is a hardware failure that will need repair.  A few things you can do before sending in for repair:

Make sure all inputs (probes and cables) are removed from all the channel inputs.

Make sure the oscilloscope has been warmed up (typically 20 minutes)

Run the oscilloscopes “Default Setup”

If not running the latest oscilloscope firmware, download and install the latest version from Tek.com.

If the oscilloscope still presents a SPC failure, the unit will need to come in for a repair


常见问题 ID 59621

查看所有常见问题 »