问题:
What is the guard terminal of an electrometer or an SMU?
答案:
The guard terminal of an electrometer will depend on the function. The guard terminal of the electrometer ammeter is the LO input terminal. The difference in potential between the HI terminal and the LO terminal is usually <1 mV. The guard terminal of the electrometer voltmeter is the preamplifier output terminal. The guard terminal of an SMU is the inside shield of the output triax connector. The SMU guard voltage is kept within 2 mV of the signal voltage on the center conductor.
此常见问题适用于:
无产品系列
产品:
常见问题 ID 71446
查看所有常见问题 »