问题:
TDS3BATB battery gauge does not display the correct charge of the battery on the TDS3000 or TDS3000B. Is the battery or the oscilloscope defective?
答案:
It is possible that neither is defective. Please try the following procedure to initialize the battery. If this does not solve the problem, please contact Tektronix service.
Recommendation:
- For all new batteries perform this procedure before putting battery into service.
- Perform this procedure on the TDS3BAT(B) every 45 days during useful service life.
TDS3BAT(B) Battery Gauge Initialization Procedure
- Completely discharge the battery through scope.
(Note: Do not connect AC during this time.) - Charge the battery fully either on the scope (TDS3000(B)) for a minimum of 36 hours first time or on the external charger (TDS3CHG) for 6 hours first time.
- Operate the oscilloscope with the battery until completely discharged and scope turns OFF. (Note: Do not connect AC during this time.)
- Recharge the battery through the oscilloscope for a minimum of 30 hours or through the external charger for a minimum of 5 hours.
- The battery gauge is now initialized and should accurately indicate the level of charge in the TDS3BATB.
- The battery gauge will indicate full when the TDS3BATB is fully charged.
(End of procedure)