问题:
I really like the idea of a removable hard disk drive for my oscilloscope+ but I’m concerned about losing the instrument calibration information. If I replace the original HDD+ will the cal constants be lost+ or is there a file that I need to transfer to the new HDD+ or are the constants stored in NVRAM?
答案:
Factory calibration data, serial number, and option keys are stored in the NVRAM.
Signal Path Compensation (SPC) data is stored on the hard drive. To be sure that your oscilloscope performance is optimized for the current temperature and to remove long-term drift, you should re-run SPC (under Utilities -> Instrument Calibration…).
此常见问题适用于:
产品系列: MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO5000B 混合信号示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 DPO7000 DPO7000 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 MSO/DPO70000DX 混合信号/数字荧光示波器 DPO7000 DPO7000