联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

Where can I get the 1740A+ 1750A or 1760 calibration software?

问题:

Where can I get the 1740A+ 1750A or 1760 calibration software?

答案:

Click the link below to download the 17xx series calibration software which normally comes on a floppy disk bundled with the service manual for these instruments. Once downloaded, unzip this archive and copy the contents of it to a floppy disk. Then run the CAL.EXE program from that floppy disk on a PC running DOS. From here the software and the 1740A service manual should walk you through the necessary steps for calibrating your instrument.

NOTE: You will need to run this software on a PC that is running true MS-DOS. Simply running the software in a DOS command windows within Windows will not work because the tool needs access to hardware ports which are protected under Windows operating systems.


此常见问题适用于:

无产品系列

产品:

常见问题 ID 62361

查看所有常见问题 »