Using TDSHT3+ my eye diagrams/ clock jitter plots are getting clipped+ or I get an empty plot. What should I do?

问题:

Using TDSHT3+ my eye diagrams/ clock jitter plots are getting clipped+ or I get an empty plot. What should I do?

答案:

Eye diagrams are expected to be around zero absolute voltage. The sink and source eye diagrams are symmetry about absolute zero. If the signal has a significant offset, there is chance of violating the top and bottom mask. If the offset is more than 75% of the Vswing the plot will be empty.

To avoid this ensure that the Probe calibration is carried over before doing the eye diagram/ Clock jitter measurement.