On the TDS1000 and TDS2000 series scopes+ where is the Trigger Point when the Pulse Width Trigger is used and a specified amount of time is selected?

问题:

On the TDS1000 and TDS2000 series scopes+ where is the Trigger Point when the Pulse Width Trigger is used and a specified amount of time is selected?

答案:

Equal: Oscilloscope triggers when trailing edge of the pulse crosses the trigger level.

Not Equal: If the pulse is narrower than the specified width, the trigger point is the trailing edge. Otherwise, the oscilloscope triggers when a pulse continues longer than the time specified as the pulse width.

Less than: The trigger point is the trailing edge.

Greater than: (also called time-out trigger): The oscilloscope triggers when a pulse continues longer than the time specified as the pulse width.