联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

Speed: How do I change the soak time for a measurement?

问题:

Speed: How do I change the soak time for a measurement?

答案:

For each measurement, a Source-Delay-Measure (SDM) sequence is executed. This implies that the output value is set (Source), the device under test is allowed to settle (Delay), and the measurement is taken (Measure). To give the device more time to settle or soak before the measurement is taken, simply change the source delay value. To change it with the front panel controls, press CONFIG I-SOURCE (or V-SOURCE) DELAY, then enter the value. Use the 'SOUR:DEL' command to accomplish the same thing over the bus.

此常见问题适用于:

产品系列: 2400 图形化系列 SMU

产品:

常见问题 ID 70891

查看所有常见问题 »