Slowest speed limitation in "internal clock?"

问题:

Slowest speed limitation in "internal clock?"

答案:

Currently, when the timebase is run at a very slow speed (multiple TDR pulses visible) in "Internal Clock" trigger setting the instrument can on rare occasions display misplaced points after the 1st square edge. This is a known problem. The impact should be minimal as this mode of operation is not useful for any particular measurement.

此常见问题适用于:

无产品系列

产品:

常见问题 ID 62106