联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

My oscilloscope has a diagnostic error+ E5111+ E5121+ E5131+ E5141+ E5151+ E5161+ E5171+ E5181+ or E5191. What does this indicate?

问题:

My oscilloscope has a diagnostic error+ E5111+ E5121+ E5131+ E5141+ E5151+ E5161+ E5171+ E5181+ or E5191. What does this indicate?

答案:

These diagnostic errors indicate a problem on the MMU (Memory Management Unit) board.

Due to the age of these products, replacement parts are no longer available. Repairs on these products are available from Tektronix Service on a best effort basis.


此常见问题适用于:

无产品系列

产品:

常见问题 ID 53006

查看所有常见问题 »