Do the signal outputs of the AFG3000 have a short circuit protection?

问题:

Do the signal outputs of the AFG3000 have a short circuit protection?

答案:

There is no need for a dedicated short-circuit protection because the output circuits are immune from a short circuit of the output. However, the outputs can be protected against over-voltage from an external source with a fuse that is included as a standard accessory.