Can the TLA7PG2 probes be hot plugged?

问题:

Can the TLA7PG2 probes be hot plugged?

答案:

The TLA7PG2 module will only recognize probes attached to the module while the TLA7PG2 application is starting up.

To prevent damage to the pattern generator module or probe, do not connect or disconnect the pattern generator cables to or from the pattern generator module or probe while the logic analyzer is powered on. The recommended DUT (Device Under Test) and pattern generator power on/off sequence is as follows: Power on the DUT first, then power on the pattern generator. Power off the pattern generator and then power off the DUT.