联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

Can I use VISA to program the 2701 over ethernet (TCP/IP)?

问题:

Can I use VISA to program the 2701 over ethernet (TCP/IP)?

答案:

Product(s) :
Integra Series: 2701

 


Question : Can I use VISA to program the 2701 over ethernet (TCP/IP)?


Answer:

In order to program the 2701 using SCPI commands sent over the VISA layer using Ethernet connection, you must add a termination character like a Line feed (LF) or a carriage return (CR).

For applications involving use of IVI drivers, you do not need to add any terminator unless you are using the low level KE2700_WriteInstrumentData function to send SCPI commands to Model 2701. With the KE2700 Write Instrument Data function you must add a terminator.

For example if using LabVIEW, you must concatenate the SCPI command string with a line feed (LF) or carriage return. This terminating character will allow the instrument to recognize that the end of the string has arrived. If using Visual Basic, you would add in a "vbcr" (Carriage Return) or a "vbLf" (Line Feed) to each SCPI command sent to the Model 2701.

 


©Copyright 2003, Keithley Instruments, Inc.


此常见问题适用于:

产品系列: Keithley 2700 万用表/数据采集/开关系统

产品:

常见问题 ID 69686

查看所有常见问题 »