联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

What happens in High Capacitance mode in Model 2450 SMU?

问题:

What happens in High Capacitance mode in Model 2450 SMU?

答案:

The SMU is a closed loop control device to maintain a steady force value on your DUT.

The capacitance of your cabling and DUT and the current sense resistor (internal to the SMU) form a pole. For certain combinations of current sense resistor and external capacitance, this can result in sufficient phase shift in the control loop to cause instability.

Enabling High C mode applies an internal capacitor across the current sense resistor to provide some phase lead to compensate for the phase lag created by capacitance of external DUT and thus restores stable control of the force value.

It has nothing to do with guard.

It has nothing to do with front panel vs. rear panel connections.

In vast majority of applications, High Capacitance mode is not needed. With 2450, you should not need to use this mode unless the load capacitance is greater than about 20nF.


此常见问题适用于:

无产品系列

产品: 2220-30-1, 2252, TPA-BNC, TPS2024

常见问题 ID 775830

查看所有常见问题 »