联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

P5200 High Voltage Differential Probe Notice

For users of:

  • TPS2000 SERIES OSCILLOSCOPES OR
  • THS700 SERIES OSCILLOSCOPES/DMM OR
  • THS3000 OSCILLOSCOPE AND
  • P5200 HIGH VOLTAGE DIFFERENTIAL PROBE

Certain of the Tektronix TPS2000 Series oscilloscopes documentation incorrectly listed the Tektronix P5200 High Voltage Differential Probe as an optional accessory. The P5200 is not compatible with the TPS2000 Series oscilloscopes.

For safe operation, do not use the Tektronix P5200 High Voltage Differential Probe with any oscilloscope or other measurement instrument that has floating inputs (isolated inputs) such as the Tektronix TPS2000 Series Oscilloscopes, Tektronix THS700 Series Oscilloscopes or THS3000 oscilloscope. The P5200 High Voltage Differential Probe requires an oscilloscope or other measurement instrument with grounded inputs. The P5200 High Voltage Differential Probe is designed to use the BNC connector ground on the product to ensure that all accessible metal on the probe remains at ground potential. The BNC connectors on the channels of any oscilloscope or other measurement instrument that has floating inputs (such as the TPS2000 Series Oscilloscopes, THS700 Series Oscilloscopes or THS3000 oscilloscope) do not provide this ground connection.

Contact Tektronix Customer Service for questions regarding this notice.