Tektronix High Voltage Probes | Oscilloscope Probes

主要特性

  • 衰减:50X/500X
  • 差动电压:500X:±1300V / 50X:±130V
  • 共模电压:±1300V

Specifications

All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.

Model Overview

Probes with the TekVPI® interface
TMDP0200 THDP0200 THDP0100
Attenuation 25X / 250X 50X / 500X 100X / 1000X
Differential Voltage (Typical)250X: ±750 V 500X: ±1500 V 1000X: ±6000 V
25X: ±75 V 50X: ±150 V 100X: ±600 V
Common Mode Voltage (Typical)±750 V±1500 V ±6000 V
Maximum Input Voltage-to-Earth (Typical)450 V CAT I1000 V CAT II2300 V CAT I
300 V CAT III600 V CAT III 1000 V CAT III
Bandwidth (Typical)200 MHz 200 MHz 100 MHz
Rise Time <2 ns <2 ns <3.6 ns
Slew Rate (Typical)<275 V/ns at 1/250 gain <650 V/ns at 1/500 gain <2500 V/ns at 1/1000 gain
Input Impedance at the Probe Tip (Typical)5 MΩ || <2 pF 10 MΩ || <2 pF 40 MΩ || <2.5 pF
Typical CMRR (Typical)DC: > 80 dBDC: > 80 dBDC: > 80 dB
100 kHz: > 60 dB100 kHz: > 60 dB100 kHz: > 60 dB
3.2 MHz: > 30 dB3.2 MHz: > 30 dB3.2 MHz: > 30 dB
100 MHz: > 26 dB100 MHz: > 26 dB100 MHz: > 26 dB
Cable Length (Typical) 1.5 m 1.5 m 1.5 m
DC Gain Accuracy±2%±2%±2%
Probes with the BNC and TekProbe® interface
P5200A P5202A P5205A P5210A
Attenuation50X / 500X 20X / 200X 50X / 500X 100X / 1000X
Differential Voltage (Typical)500X: ±1300 V200X: ±640 V500X: ±1300 V1000X: ±5600 V
50X: ±130 V20X: ±64 V50X: ±130 V100X: ±560 V
Common Mode Voltage (Typical)±1300 V ±640 V ±1300 V ±5600 V
Maximum Input Voltage-to-Earth (Typical)1000 V CAT II 450 V CAT l

1000 V CAT II

2300 V CAT l
300 V CAT ll 1000 V CAT III
Bandwidth (Typical)50 MHz 100 MHz 100 MHz 50 MHz
Differential Input Impedance (Typical)10 MΩ || 2 pF 5 MΩ || 2 pF 10 MΩ || 2 pF 40 MΩ || 2.5 pF
Input Impedance between each Input and Ground (Typical)5 MΩ || 4 pF 2.5 MΩ || 4 pF 5 MΩ || 4 pF 20 MΩ || 5 pF
Typical CMRR (Typical)DC: >80 dBDC: >80 dB
100 kHz: >60 dB100 kHz: >60 dB
3.2 MHz: >30 dB3.2 MHz: >40 dB
50 MHz: >26 dB50 MHz: >30 dB
Cable length (Typical)1.5 m
DC Gain Accuracy±2%±2%±2%±2%

THDP0200 probe label

TMDP P5200Datasheet 51W1119515

1. The differential voltages the maximum measurable range between the (+) and (-) input leads of the probe. Beyond these limits, the output could be clipped.

2. The maximum common mode voltage and maximum input voltage-to-earth (RMS) are the maximum voltages that each input lead (+/-) can be from ground.