Document #: SPEC-2410E
SOURCE SPECIFICATIONS1
VOLTAGE PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE | PROGRAMMING RESOLUTION | ACCURACY (1 Year) 23°C ±5°C ±(% rdg. + volts) | NOISE (peak-peak) 0.1Hz – 10Hz |
200.00 mV | 5 µV | 0.02% + 600 µV | 5 µV |
2.00000 V | 50 µV | 0.02% + 600 µV | 50 µV |
20.0000 V | 500 µV | 0.02% + 2.4 mV | 5 mV |
1000.00 V | 50 mV | 0.02% + 100 mV | 20 mV |
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C): ±(0.15 × accuracy specification)/ °C.
MAX. OUTPUT POWER: 22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS: ±21V @ ±1.05A, ±1100V @ ±21mA.
VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 1mV.
NOISE 10Hz – 1MHz (p-p): 20mV typical into a resistive load.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = none.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (full scale step, resistive load, 20mA range).
CURRENT PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE | PROGRAMMING RESOLUTION | ACCURACY (1 Year)3 23°C ±5°C ±(% rdg. + volts) | NOISE (peak-peak) 0.1Hz – 10Hz |
1.00000 µA | 50 pA | 0.035% + 600 pA | 5 pA |
10.0000 µA | 500 pA | 0.033% + 2 nA | 50 pA |
100.000 µA | 5 nA | 0.031% + 20 nA | 500 pA |
1.00000 mA | 50 nA | 0.034% + 200 nA | 5 nA |
20.0000 mA | 500 nA | 0.045% + 4 µA | 200 nA |
100.000 mA | 5 µA | 0.066% + 20 µA | 1 µA |
1.00000 A2 | 50 µA | 0.27% + 900 µA | 100 µA |
TEMPERATURE COEFFICIENT (0° to 18°C and 28° to 50°C): ±(0.15 × accuracy specification)/ °C.
MAX. OUTPUT POWER: 22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS: ±1.05A @ ±21V, ±21 mA @ ±1100V.
CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range + 1nA.
VOLTAGE LIMIT: Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (1mA step, RL = 10kΩ, 20V range).
ADDITIONAL SOURCE SPECIFICATIONS
TRANSIENT RESPONSE TIME: 30µs minimum for the output to recover to its specification following a step change in load.
COMMAND PROCESSING TIME: Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent < nrf > command.
Autorange On: 10ms. Autorange Off: 7ms.
OUTPUT SETTLING TIME: Time required to reach 0.1% of final value after command is processed. 100µs typical. Resistive load. 10µA to 100mA range.
OUTPUT SLEW RATE (±30%): 0.5V/µs, 1000V range, 20mA compliance. 0.15V/µs, 20V range, 100mA compliance.
DC FLOATING VOLTAGE: Output can be floated up to ±250VDC from chassis ground.
REMOTE SENSE: Up to 1V drop per load lead.
COMPLIANCE ACCURACY: Add 0.3% of range and ±0.02% of reading to base specification.
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT: Overshoot into a fully resistive 100kΩ load, 10Hz to 1MHz BW, adjacent range changes between 200mV, 2V, and 20V ranges, 100mV typical.
MINIMUM COMPLIANCE VALUE: 0.1% of range.
1. Specifications valid for continuous output currents below 105mA. For operation above 105mA continuous for > 1 minute, derate accuracy 10%/35mA above 105mA.
2. Full operation (1A) regardless of load to 30°C. Above 30°C ambient, derate 35mA/°C and prorate 35mA/Ω load. 4-wire mode. For current sink operation on 1A range, maximum continuous power is limited to approximately 1/2 rated power or less, depending on current, up to 30°C ambient. See power equations in the User’s Manual to calculate allowable duty cycle for specific conditions.
3. For sink mode, 1µA to 100mA range, accuracy is ±(0.5% + offset*3). For 1A range, accuracy is: ±(0.15 × accuracy specification)/ °C on 20 V and 1000 V ranges ±(1.5% + offset*3)
MEASURE SPECIFICATIONS1,2
VOLTAGE MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE | DEFAULT RESOLUTION | INPUT RESISTANCE | ACCURACY (1 Year) 23°C ±5°C ±(% rdg. + volts) |
200.00 mV | 1 µV | >10 GΩ | 0.012% + 300 µV |
2.00000 V | 10 µV | >10 GΩ | 0.012% + 300 µV |
20.0000 V | 100 µV | >10 GΩ | 0.015% + 1 mV |
1000.00 V | 10 mV | >10 GΩ | 0.015% + 50 mV |
TEMPERATURE COEFFICIENT (0° to 18°C and 28° to 50°C): ±(0.15 × accuracy specification)/ °C on 20 V and 1000 V ranges
±(0.30 × accuracy specification)/ °C on 200 mV and 2 V ranges
CURRENT MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE | DEFAULT RESOLUTION | VOLTAGE BURDEN3 | ACCURACY (1 Year) 23°C ±5°C ±(% rdg. + amps) |
1.00000 µA | 10 pA | <1 mV | 0.029% + 300 pA |
10.0000 µA | 100 pA | <1 mV | 0.027% + 700 pA |
100.000 µA | 1 nA | <1 mV | 0.025% + 6 nA |
1.00000 mA | 10 nA | <1 mV | 0.027% + 60 nA |
20.0000 mA | 100 nA | <1 mV | 0.035% + 1.2 µA |
100.000 mA | 1 µA | <1 mV | 0.055% + 6 µA |
1.00000 A | 10 µA | <1 mV | 0.22% + 570 µA |
TEMPERATURE COEFFICIENT (0° to 18°C and 28° to 50°C): ±(0.10 × accuracy specification)/ °C.
RESISTANCE MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE | DEFAULT RESOLUTION | DEFAULT TEST CURRENT | NORMAL ACCURACY (23°C ±5°C) 1 YEAR, ±(% rdg. + ohms) |
<2.00000 Ω4 | - | - | Source IACC + Meas. VACC |
20.0000 Ω | 100 µΩ | 100 mA | 0.11% + 0.006 Ω |
200.000 Ω | 1 mΩ | 10 mA | 0.09% + 0.12 Ω |
2.00000 kΩ | 10 mΩ | 1 mA | 0.08% + 0.6 Ω |
20.0000 kΩ | 100 mΩ | 100 µA | 0.07% + 6 Ω |
200.000 kΩ | 1 Ω | 10 µA | 0.07% + 60 Ω |
2.00000 MΩ | 10 Ω | 1 µA | 0.12% + 600 Ω |
20.0000 MΩ | 100 Ω | 1 µA | 0.12% + 2.4 kΩ |
200.000 MΩ | 1 kΩ | 100 nA | 0.66% + 24 kΩ |
>200.000 MΩ4 | - | - | Source IACC + Meas. VACC |
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C): ±(0.15 × accuracy specification)/ °C.
SOURCE I MODE, MANUAL OHMS: Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense).
SOURCE V MODE, MANUAL OHMS: Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense).
6-WIRE OHMS MODE: Available using active ohms guard and guard sense (except on 1A and 1000V ranges). Max. Guard Output Current: 40mA typical. Accuracy is load dependent. Refer to White Paper no. 2033 for calculation formula.
GUARD OUTPUT IMPEDANCE: <0.1Ω in ohms mode.
1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV and 1A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, and 1A ranges, add 0.5%.
2. Accuracies apply to 2- or 4-wire mode when properly zeroed.
3. 4-wire mode.
4. Manual ohms only.
CONTACT CHECK SPECIFICATIONS
SPEED: 350µs for verification and notification.
CONTACT CHECK: | 2Ω | 15Ω | 50Ω |
No contact check failure | <1.00Ω | <13.5Ω | <47.5Ω |
Always contact check failure | >3.00Ω | >16.5Ω | >52.5Ω |
SYSTEM SPEEDS
MEASUREMENT1
MAXIMUM RANGE CHANGE RATE: 70/second.
MAXIMUM MEASURE AUTORANGE TIME: 40ms (fixed source)2.
SWEEP OPERATION3 READING RATES (rdg./second) FOR 60Hz (50Hz):
SPEED | NPLC/TRIGGER ORIGIN | MEASURE | SOURCE-MEASURE5 | SOURCE-MEASURE PASS/FAIL TEST4,5 | SOURCE-MEMORY4,5 | ||||
TO MEM. | TO GPIB | TO MEM | TO GPIB | TO MEM. | TO GPIB | TO MEM. | TO GPIB | ||
Fast | 0.01 / internal | 2081 (2030) | 1754 | 1551 (1515) | 1369 | 902 (900) | 981 | 165 (162) | 165 |
IEEE-488.1 Mode | 0.01 / external | 1239 (1200) | 1254 | 1018 (990) | 1035 | 830 (830) | 886 | 163 (160) | 163 |
Fast | 0.01 / internal | 2801 (2030) | 1198 (1210) | 1551 (1515) | 1000 (900) | 902 (900) | 809 (840) | 165 (162) | 164 (162) |
IEEE-488.2 Mode | 0.01 / external | 1239 (1200) | 1079 (1050) | 1018 (990) | 916 (835) | 830 (830) | 756 (780) | 163 (160) | 162 (160) |
Medium | 0.10 / internal | 510 (433) | 509 (433) | 470 (405) | 470 (410) | 389 (343) | 388 (343) | 133 (126) | 132 (126) |
IEEE-488.2 Mode | 0.10 / external | 438 (380) | 438 (380) | 409 (360) | 409 (365) | 374 (333) | 374 (333) | 131 (125) | 131 (125) |
Normal | 1.00 / internal | 59 (49) | 59 (49) | 58 (48) | 58 (48) | 56 (47) | 56 (47) | 44 (38) | 44 (38) |
IEEE-488.2 Mode | 1.00 / external | 57 (48) | 57 (48) | 57 (48) | 57 (47) | 56 (47) | 56 (47) | 44 (38) | 44 (38) |
SINGLE READING OPERATION READING RATES (rdg./second) FOR 60Hz (50Hz):
SPEED | NPLC/TRIGGER ORIGIN | MEASURE TO GPIB | SOURCE-MEASURE TO GPIB5 | SOURCE-MEASURE PASS/FAIL TEST4,5 TO GPIB |
Fast (488.1) | 0.01 / internal | 537 | 140 | 135 |
Fast (488.2) | 0.01 / internal | 256 (256) | 79 (83) | 79 (83) |
Medium (488.2) | 0.10 / internal | 167 (166) | 72 (70) | 69 (70) |
Normal (488.2) | 1.00 / internal | 49 (42) | 34 (31) | 35 (30) |
COMPONENT INTERFACE HANDLER TIME FOR 60Hz (50Hz):4,6
SPEED | NPLC/TRIGGER ORIGIN | MEASURE TO GPIB | SOURCE PASS/FAIL TEST | SOURCE-MEASURE PASS/FAIL TEST4,5 TO GPIB |
Fast | 0.01/ external | 1.04 ms (1.08 ms) | 0.5 ms (0.5 ms) | 4.82 ms (5.3 ms) |
Medium | 0.10 / external | 2.55 ms (2.9 ms) | 0.5 ms (0.5 ms) | 6.27 ms (7.1 ms) |
Normal | 1.00 / external | 17.53 ms (20.9 ms) | 0.5 ms (0.5 ms) | 21.31 ms (25.0 ms) |
1. Reading rates applicable for voltage or current measurements. Auto zero off, auto range off, filter off, display off, trigger delay = 0, source auto-clear off, and binary reading format.
2. Purely resistive load. 1µA and 10µA ranges <65ms.
3. 1000 point sweep was characterized with the source on a fixed range.
4. Pass/Fail test performed using one high limit and one low math limit.
5. Includes time to re-program source to a new level before making measurement.
6. Time from falling edge of START OF TEST signal to falling edge of END OF TEST signal.
7. Command processing time of : SOURce:VOLTage|CURRent:TRIGgered < nrf >
command not included.
General
NOISE REJECTION:
NPLC | NMRR | CMRR | |
Fast | 0.01 | - | 80 dB |
Medium | 0.1 | - | 80 dB |
Slow | 1 | 60 dB | 100 dB1 |
1. Except lowest 2 current ranges – 90dB.
LOAD IMPEDANCE: Stable into 20,000pF typical.
COMMON MODE VOLTAGE: 250 VDC.
COMMON MODE ISOLATION: >109Ω, <1000pF.
OVERRANGE: 105% of range, source and measure.
MAX. VOLTAGE DROP BETWEEN INPUT/OUTPUT AND SENSE TERMINALS: 5V.
MAX. SENSE LEAD RESISTANCE: 50kΩ for rated accuracy.
SENSE INPUT IMPEDANCE: >1010Ω.
GUARD OFFSET VOLTAGE: <300µV, typical.
SOURCE OUTPUT MODES:
Fixed DC level
Memory List (mixed function)
Stair (linear and log)
SOURCE MEMORY LIST: 100 points max.
MEMORY BUFFER: 5,000 readings @ 5.5 digits (two 2,500 point buffers).
Includes selected measured values and timestamp.
Lithium battery backup (3 yr+ battery life).
PROGRAMMABILITY: IEEE-488 (SCPI-1996.0), RS-232, 5 user-definable power-up states plus factory default and *RST.
DIGITAL INTERFACE:
Output Enable: Active low input.
Handler Interface: Start of test, end of test, 3 category bits. +5V @ 300mA supply.
Digital I/O: 1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode clamped).
POWER SUPPLY: 100V to 240V rms, 50-60Hz (automatically detected at power up). 210VA.
COOLING: Forced air, variable speed.
WARRANTY: 1 year.
EMC: Conforms to European Union Directive 89/336/EEC, EN 61326-1.
SAFETY: Conforms to European Union Directive 73/23/EEC, EN61010-1.
WARM-UP: 1 hour to rated accuracies.
DIMENSIONS: 89mm high × 213mm wide × 370mm deep (3 1/2 in × 8 3/8 in × 14 9/16 in). Bench Configuration (with handle & feet):104mm high × 238mm wide × 370mm deep (4 1/8 in × 9 3/8 in × 14 9/16 in).
WEIGHT: 3.3kg (7.3 lb).
ENVIRONMENT:
For Indoor Use Only: Maximum 2000m above Sea Level
Operating: 0°–50°C, 70%R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage: –25°C to 65°C.
Specifications subject to change without notice.
SourceMeter® is a registered trademark of Keithley Instruments, Inc. All other trademarks are the property of their respective companies.
Find more valuable resources at TEK.COM
Copyright © Tektronix. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.
HW 3/20/03 Rev. D