Document #SPEC-2000M
Specifications are subject to change without notice.
DC CHARACTERISTICS
CONDITIONS: MED (1 PLC)1 or SLOW (10 PLC) or MED (1 PLC) with filter of 10
ACCURACY: ±(ppm of reading + ppm of range) (ppm = parts per million) (e.g., 10ppm = 0.001%)
FUNCTION | RANGE | RESOLUTION | TEST CURRENT OR BURDEN VOLTAGE (±5%) | Input Resistance | 24 HOUR14 23°C ± 1° | 90 DAY 23°C ± 5° | 1 YEAR 23°C ± 5° | TEMPERATURE COEFFICIENT 0°–18°C & 28°–50°C |
Voltage | 100.0000 mV | 0.1 µV | < 10 GΩ | 30 + 30 | 40 + 35 | 50 + 35 | 2 + 6 | |
1.000000 V | 1.0 µV | < 10 GΩ | 15 + 6 | 25 + 7 | 30 + 7 | 2 + 1 | ||
10.00000 V | 10 µV | < 10 GΩ | 15 + 4 | 20 + 5 | 30 + 5 | 2 + 1 | ||
100.0000 V | 100 µV | 10 MΩ ±1% | 15 + 6 | 30 + 6 | 45 + 6 | 5 + 1 | ||
1000.000 V9 | 1 mV | 10 MΩ ±1% | 20 + 6 | 35 + 6 | 45 + | 5 + 1 | ||
Resistance15 | 100.0000 Ω | 100 µΩ | 1 mA | 30 + 30 | 80 + 40 | 100 + 40 | 8 + 6 | |
1.000000 kΩ | 1 mΩ | 1 mA | 20 + 6 | 80 + 10 | 100 + 10 | 8 + 1 | ||
10.00000 kΩ | 10 mΩ | 100 µA | 20 + 6 | 80 + 10 | 100 + 10 | 8 + 1 | ||
100.0000 kΩ | 100 mΩ | 7 µA | 20 + 6 | 80 + 10 | 100 + 10 | 8 + 1 | ||
1.000000 MΩ16 | 1 Ω | 7 µA | 20 + 6 | 80 + 10 | 100 + 10 | 8 + 1 | ||
10.00000 MΩ11,16 | 10 Ω | 700 nA // 10MΩ | 150 + 6 | 200 + 10 | 400 + 10 | 95 + 1 | ||
100.0000 MΩ11,16 | 100 Ω | 700 nA // 10MΩ | 800 + 30 | 1500 + 30 | 1500 + 30 | 900 + 1 | ||
Current | 10.00000 mA | 10 nA | < 0.15 V | 60 + 30 | 300 + 80 | 500 + 80 | 50 + 5 | |
100.0000 mA | 100 nA | < 0.03 V | 100 + 300 | 300 + 800 | 500 + 800 | 50 + 50 | ||
1.000000 A | 1 µA | < 0.3 V | 200 + 30 | 500 + 80 | 800 + 80 | 50 + 5 | ||
3.00000 A | 10 µA | < 1 V | 1000 + 15 | 1200 + 40 | 1200 + 40 | 50 + 5 | ||
Continuity 2W | 1 kΩ | 100 mΩ | 1 mA | 40 + 100 | 100 + 100 | 120 + 100 | 8 + 1 | |
Diode Test | 3.00000 V | 10 µV | 1 mA | 20 + 6 | 30 + 7 | 40 + 7 | 8 + 1 | |
10.00000 V | 10 µV | 70 µA | 20 + 6 | 30 + 7 | 40 + 7 | 8 + 1 | ||
10.00000 V | 10 µV | 7 µA | 20 + 6 | 30 + 7 | 40 + 7 | 8 + 1 |
DC OPERATING CHARACTERISTICS2
FUNCTION | DIGITS | READINGS/s | PLCs 8 |
DCV (all ranges), DCI (all ranges), and Ohms (<10M range) | 6.53,4 | 5 | 10 |
6.53,7 | 30 | 1 | |
6.53,5 | 50 | 1 | |
5.53,5 | 270 | 0.1 | |
5.55 | 500 | 0.1 | |
5.55 | 1000 | 0.04 | |
4.55 | 2000 | 0.01 |
DC SYSTEM SPEEDS2, 6
RANGE CHANGE3: 50/s.
FUNCTION CHANGE3: 45/s.
AUTORANGE TIME3, 10: <30 ms.
ASCII READINGS TO RS-232 (19.2K BAUD): 55/s.
MAX. INTERNAL TRIGGER RATE: 2000/s.
MAX. EXTERNAL TRIGGER RATE: 500/s.
DC GENERAL
LINEARITY OF 10VDC RANGE: ±(1ppm of reading + 2ppm of range).
DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION: 1000V, all ranges.
MAXIMUM 4WΩ LEAD RESISTANCE: 10% of range per lead for 100Ω and 1kΩ ranges; 1kΩ per lead for all other ranges.
DC CURRENT INPUT PROTECTION: 3A, 250V fuse.
SHUNT RESISTOR: 0.1Ω for 3A, 1A, and 100mA ranges. 10Ω for 10mA range.
CONTINUITY THRESHOLD: Adjustable 1Ω to 1000Ω.
AUTOZERO OFF ERROR: Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change.
OVERRANGE: 120% of range except on 1000V, 3A and Diode
SPEED AND NOISE REJECTION
RATE | READINGS/S | DIGITS | RMS NOISE 10V RANGE | NMRR12 | CMRR13 |
10 PLC | 5 | 6.5 | < 1.5 µV | 60 dB | 140 dB |
1 PLC | 50 | 6.5 | < 4 µV | 60 dB | 140 dB |
0.1 PLC | 500 | 5.5 | < 22 µV | — | 80 dB |
0.01 PLC | 2000 | 4.5 | < 150 µV | — | 80 dB |
DC Notes
1 Add the following to “ppm of range” uncertainty:1V and 100V, 2ppm; 100mV, 15ppm; 100Ω, 15ppm; 1K-1MΩ, 2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm.
2 Speeds are for 60Hz operation using factory default operating conditions (*RST). Autorange off, Display off, Trigger delay= 0.
3 Speeds include measurement and binary data transfer out the GPIB.
4 Auto zero off.
5 Sample count = 1024, auto zero off.
6 Auto zero off, NPLC = 0.01.
7 Ohms = 24 readings/second.
8 1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up.
9 For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
10 Add 120ms for ohms.
11 Must have 10% matching of lead resistance in Input HI and LO.
12 For line frequency ±0.1%.
13 For 1kΩ unbalance in LO lead.
14 Relative to calibration accuracy.
15 Specifications are for 4-wire ohms. For 2-wire ohms, add 1Ω additional uncertainty.
16 For rear inputs, add the following to Temperature Coefficient “ppm of reading” uncertainty: 10MΩ 70ppm, 100MΩ 385ppm. Operating environment specified for 0° to 50°C and 50% RH at 35°C.
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY 1: ±(% of reading + % of range), 23°C ±5 °C
VOLTAGE RANGE | RESOLUTION | CALIBRATION CYCLE | 3 Hz–10 Hz10 | 10 Hz–20 kHz | 20 kHz–50 kHz | 50 kHz–100 kHz | 100 kHz–300 kHz |
100.0000 mV | 0.1 µV | ||||||
1.000000 V | 1.0 µV | 90 Days | 0.35 + 0.03 | 0.05 + 0.03 | 0.11 + 0.05 | 0.60 + 0.08 | 4 + 0.5 |
10.00000 V | 10 µV | ||||||
100.0000 V | 100 µV | 1 Year | 0.35 + 0.03 | 0.06 + 0.03 | 0.12 + 0.05 | 0.60 + 0.08 | 4 + 0.5 |
750.000 V | 1 mV | ||||||
TEMPERATURE COEFFICIENT/°C8 | 0.035 + 0.003 | 0.005 + 0.003 | 0.006 + 0.005 | 0.01 + 0.006 | 0.03 + 0.01 |
CURRENT RANGE | RESOLUTION | CALIBRATION CYCLE | 3 Hz–10 Hz | 10 Hz–3 kHz | 3 kHz–5 kHz |
1.000000 A | 1 µA | 90 Day/1 Year | 0.30 + 0.04 | 0.10 + 0.04 | 0.14 + 0.04 |
3.000009 A | 10 µA | 90 Day/1 Year | 0.35 + 0.06 | 0.15 + 0.06 | 0.18 + 0.06 |
TEMPERATURE COEFFICIENT/°C8 | 0.035 + 0.006 | 0.015 + 0.006 | 0.015 + 0.006 |
HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading)7
CREST FACTOR: | 1-2 | 2-3 | 3-4 | 4-5 |
ADDITIONAL ERROR: | 0.05 | 0.15 | 0.30 | 0.40 |
AC OPERATING CHARACTERISTICS2
FUNCTION | DIGITS | READINGS/s | RATE | BANDWIDTH |
ACV (all ranges), and ACI (all ranges) | 6.53 | 2s/reading | SLOW | 3 Hz–300 kHz |
6.53 | 1.4 | MED | 30 Hz–300 kHz | |
6.54 | 4.8 | MED | 30 Hz–300 kHz | |
6.53 | 2.2 | FAST | 300 Hz–300 kHz | |
6.54 | 35 | FAST | 300 Hz–300 kHz |
ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading)
SLOW | MED | FAST | |
20Hz – 30Hz | 0 | 0.3 | — |
30Hz – 50Hz | 0 | 0 | — |
50Hz – 100Hz | 0 | 0 | 1.0 |
100Hz – 200Hz | 0 | 0 | 0.18 |
200Hz – 300Hz | 0 | 0 | 0.10 |
>300Hz | 0 | 0 | 0 |
AC SYSTEM SPEEDS2, 5
FUNCTION/RANGE CHANGE6 4/s.
AUTORANGE TIME: <3s.
ASCII READINGS TO RS-232 (19.2k BAUD)4: 50/s.
MAX. INTERNAL TRIGGER RATE4: 300/s.
MAX. EXTERNAL TRIGGER RATE4: 300/s.
AC GENERAL
INPUT IMPEDANCE: 1MΩ ±2% paralleled by <100pF.
ACV INPUT PROTECTION: 1000Vp.
MAXIMUM DCV: 400V on any ACV range.
ACI INPUT PROTECTION: 3A, 250V fuse.
BURDEN VOLTAGE: 1A Range: <0.3V rms. 3A Range: <1V rms.
SHUNT RESISTOR: 0.1Ω on all ACI ranges.
AC CMRR: >70dB with 1kΩ in LO lead.
MAXIMUM CREST FACTOR: 5 at full scale.
VOLT HERTZ PRODUCT: ≤8 x 107 V·Hz.
OVERRANGE: 120% of range except on 750V and 3A ranges.
AC Notes
1 Specifications are for SLOW rate and sinewave inputs >5% of range.
2 Speeds are for 60 Hz operation using factory default operating conditions (*RST). Auto zero off, Auto range off, Display off, includes measurement and binary data transfer out the GPIB.
3 0.01% of step settling error. Trigger delay = 400ms.
4 Trigger delay = 0.
5 DETector:BANDwidth 300, NPLC = 0.01.
6 Maximum useful limit with trigger delay = 175ms.
7 Applies to non-sinewaves >5Hz and <500Hz. (Guaranteed by design for Crest Factors >4.3)
8 Applies to 0°–18°C and 28°–50°C.
9 For signal levels > 2.2A, add additional 0.4% to “of reading” uncertainty.
10 Typical uncertainties. Typical represents two sigma or 95% of manufactured units measure < 0.35% of reading and three sigma or 99.7% < 1.06% of reading.
FREQUENCY AND PERIOD CHARACTERISTICS1,2
ACV RANGE | FREQUENCY RANGE | PERIOD RANGE | GATE TIME | RESOLUTION ±(ppm of reading) | ACCURACY 90 DAY/1 YEAR ±(% of reading) |
100 mV to 750 V | 3 Hz to 500 kHz | 333 ms to 2 µs | 1 s (SLOW) | 0.3 | 0.01 |
Frequency Notes
1 Specifications are for square wave inputs only. Input signal must be >10% of ACV range. If input is <20mV on the 100mV range then frequency must be >10Hz.
2 20% overrange on all ranges except 750V range.
TEMPERATURE CHARACTERISTICS
THERMOCOUPLE2, 3, 4 | 90 DAY/1 YEAR (23°C ± 5°C) ACCURACY1 | |||
TYPE | RANGE | RESOLUTION | Relative to Reference Junction | Using5 2001-TCSCAN |
J | –200 to + 760°C | 0.001°C | ±0.5°C | ±0.65°C |
K | –200 to + 1372°C | 0.001°C | ±0.5°C | ±0.70°C |
T | –200 to + 400°C | 0.001°C | ±0.5°C | ±0.68°C |
Temperature Notes
1 For temperatures <–100°C, add ±0.1°C and >900°C add ±0.3°C.
2 Temperature can be displayed in °C, K or °F.
3 Accuracy based on ITS-90.
4 Exclusive of thermocouple error.
5 Specifications apply to channels 2-6. Add 0.06°C/channel from channel 6.
INTERNAL SCANNER SPEED4
MAXIMUM INTERNAL SCANNER RATES
RANGE: Channels/s1
TRIGGER DELAY = 0
DCV2 | ACV2, 3 | 2-WIRE OHMS2 | 4-WIRE OHMS2 | TEMPERATURE 2 |
All : 110 | All : 100 | All : 105 | <10MΩ : 33 | All : 60 |
TRIGGER DELAY = AUTO
DCV2 | ACV2, 3 | 2-WIRE OHMS2 | 4-WIRE OHMS2 | TEMPERATURE2 |
0.1 V : 105 | All : 1.8 | 100 Ω : 85 | 100 Ω : 29 | All : 60 |
1 V : 105 | 1 kΩ : 85 | 1 kΩ : 29 | ||
10 V : 105 | 10 kΩ : 42 | 10 kΩ : 22 | ||
100 V : 70 | 100 kΩ : 28 | 100 kΩ : 18 | ||
1000 V : 70 | 1 MΩ : 8 | 1 MΩ : 7 | ||
10 MΩ : 5 | 10 MΩ : 5 | |||
100 MΩ : 3 | 100 MΩ : 3 |
Internal Scanner Speed Notes
1 Speeds are for 60Hz operation using factory default operating conditions (*RST). Auto Zero off, Auto Range off, Display off, sample count = 1024.
2 NPLC = 0.01.
3 DETector:BANDwidth 300.
4 10-channel card specification. See individual card specifications for options other than 10-channel card.
TRIGGERING AND MEMORY
READING HOLD SENSITIVITY: 0.01%, 0.1%, 1%, or 10% of reading.
TRIGGER DELAY: 0 to 99 hrs (1ms step size).
EXTERNAL TRIGGER LATENCY: 200µs + <300µs jitter with autozero off, trigger delay = 0.
MEMORY: 1024 readings.
MATH FUNCTIONS
Rel, Min/Max/Average/StdDev (of stored reading), dB, dBm, Limit Test, %, and mX+b with user defined units displayed.
dBm REFERENCE RESISTANCES: 1 to 9999Ω in 1Ω increments.
STANDARD PROGRAMMING LANGUAGES
SCPI (Standard Commands for Programmable Instruments)
Keithley 196/199
Fluke 8840A. Fluke 8842A
REMOTE INTERFACE
GPIB (IEEE-488.1, IEEE-488.2) and RS-232C.
GENERAL
POWER SUPPLY: 100V / 120V / 220V / 240V.
LINE FREQUENCY: 50Hz to 60Hz and 400Hz, automatically sensed at power-up.
POWER CONSUMPTION: 22 VA.
OPERATING ENVIRONMENT: Specified for 0°C to 50°C. Specified to 80% R.H. at 35°C. Altitude up to 2000 meters.
STORAGE ENVIRONMENT: –40°C to 70°C.
WARRANTY: 1 year.
EMC: Conforms to European Union EMC Directive.
SAFETY: Conforms to European Union Low Voltage Directive.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARMUP: 1 hour to rated accuracy.
DIMENSIONS:
- Rack Mounting: 89mm high x 213mm wide x 370mm deep (3.5 in. x 8.38 in. x 14.56 in.).
- Bench Configuration (with handle and feet): 104mm high x 238mm wide x 370mm deep (4.13 in. x 9.38 in. x 14.56 in.).
NET WEIGHT: 2.9kg (6.3 lb).
SHIPPING WEIGHT: 5kg (11 lb).
VOLT HERTZ PRODUCT: ≤8 x 107V·Hz.